Magnetic Stirrer Hotplate Stirrer

A magnetic stirrer drives a PTFE-encapsulated stir bar via a rotating magnet beneath the base plate, coupling torque through the vessel wall without shaft penetration. Combined hotplate models integrate resistance heating for simultaneous temperature control. Processing 1 mL to 5 L at viscosities up to approximately 1,000 cP, it provides simple, contamination-free stirring for dissolution studies, chemical reactions, and analytical sample preparation.

Challenges

Stir Bar Decoupling at High Viscosity or Speed

When fluid resistance exceeds the magnetic coupling torque, the stir bar loses synchrony with the drive magnet, spinning erratically or stopping entirely, giving a false impression of active mixing.

Inadequate Mixing in Tall or Large Vessels

Magnetic stirrers generate primarily bottom-level circulation; in vessels taller than approximately twice the vessel diameter, upper liquid layers receive little mixing.

Hotplate Temperature Overshoot

Without a probe directly in the liquid, the plate temperature can significantly overshoot the target, risking thermal degradation of heat-sensitive samples.

Stir Bar Retrieval and Cross-Contamination

In multi-product laboratory environments, stir bars are frequently misplaced, damaged, or inadequately cleaned between uses, creating cross-contamination risk.

Ineffective Solid Suspension

Denser solid particles settle faster than the toroidal flow can lift them, leaving undissolved solids resting beneath or beside the stir bar rather than suspended in the bulk.

Solutions

Applications

Models & Capacities

Standard Hotplate Magnetic Stirrer — Models & Capacities (Micro 1 mL to Pilot 50 L)

Model

Scale

Working Vol. [mL – L]

Plate / Bath Size [mm]

Speed Range [RPM]

Heating Power [W]

Max Plate Temp [°C]

Temp Accuracy

Temp Sensor

Max Viscosity [cP]

Display / Control

Overall W×D×H [mm]

Weight [kg]

 

MS-HP-MICRO

Micro/Lab

1–50 mL

80×80

200–2,000

10

200

±2°C (plate)

Pt 1000

<200

Analogue knob

110×110×75

0.6

 

MS-HP-MINI

Lab

1–100 mL

100×100

100–2,000

15

320

±1°C (plate)

Pt 1000

<500

Analogue knob

130×130×85

0.8

 

MS-HP-100

Lab

20–250 mL

120×120

50–1,800

20

340

±1°C (plate)

Pt 1000

<500

Analogue knob

155×155×90

1.0

 

MS-HP-500A

Lab

50 mL–500 mL

130×130

50–1,600

30

350

±1°C (plate)

Pt 1000

<1,000

Analogue

    160×160×95

1.2

 

MS-HP-1L-A

Lab

50 mL–1 L

140×140

50–1,500

50

360

±1°C (plate)

Pt 1000

<2,000

Analogue

175×175×100

1.5

 

MS-HP-1L-D

Lab

50 mL–1 L

140×140

50–1,500

50

360

±0.5°C (probe)

Pt 100 ext.

<2,000

Digital LED

175×175×105

1.8

 

MS-HP-2L

Lab

100 mL–2 L

155×155

50–1,500

100

380

±0.5°C (probe)

Pt 100 ext.

<5,000

Digital LED

200×200×110

2.2

 

MS-HP-3L

Lab/Pilot

250 mL–3 L

170×170

50–1,400

150

380

±0.5°C (probe)

Pt 100 ext.

<8,000

Digital LED

215×215×115

2.8

 

MS-HP-5L

Lab/Pilot

250 mL–5 L

180×180

50–1,200

200

400

±0.5°C (probe)

Pt 100 ext.

<10,000

Digital LED

230×230×120

3.2

 

MS-HP-10L

Lab/Pilot

1–10 L

220×220

30–1,000

300

380

±0.5°C (probe)

Pt 100 ext.

<15,000

Digital

    275×275×130

4.5

 

MS-HP-20L

Pilot

5–20 L

280×280

20–800

450

350

±1°C (probe)

Pt 100 ext.

<20,000

Digital PID

340×340×145

7.0

 

MS-HP-50L

Pilot

10–50 L

380×380

10–600

800

320

±1°C (probe)

Pt 100 ext.

<20,000

Digital PID

450×450×160

12

 

 

×